LOGIQ 400 SERVICE MANUAL
FUNCTIONAL CHECKS
4–45
REV 6
4–4–2 Power Supply Adjustment Procedure (continued)
POWER SUPPLIES ON LV3 UNIT
ILLUSTRATION 4–33
Note
Check the +48V output as necessary. Refer to ILLUSTRATION 4–33 for the measuring points.
Torelance: +48V ±2.4V
Note
Use a non–conductive trim–pot to adjust the power supply.