LOGIQ 400 SERVICE MANUALGE MEDICAL SYSTEMS
2127661
RENEWAL PARTS
6–160
REV 9
6–2–57 Tracball Assy (FRU No. 457)
Time Required
20 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–80.
1. Turn OFF the system.
2. Remove the Keyboard Panel Assy. Refer to ILLUSTRATION 6–77 on page 6–155.
3. Unscrew four screws (
1
3
).
4. Disconnect one connector (
4
).
5. Unscrew four screws (
5
8
).
6. Remove The Holding Metal Plate.
TRACKBALL DISASSEMBLY
ILLUSTRATION 6–80

Table of Contents

Other manuals for GE LOGIQ 400

Related product manuals