EasyManua.ls Logo

GE LOGIQ 400 - Page 334

GE LOGIQ 400
526 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–204
REV 9
6–2–81 Time Keeper RAM (for MVME167–002B) (FRU No. 613C)
Note
The MVME167–002B is supplied with the optional DICOM option.
Time Required
1 Hour
Tool Required
Screwdriver
IC Remover
Procedure
Refer to ILLUSTRATION 6–116 on page 6–205.
1. Turn OFF the system.
2. Remove the Side Cover Left (FRU 301). Refer to 6–2–31 on page 6–127.
3. Unscrew the four Nest Cover screws.
4. Remove the Nest Cover.
5. Remove the CPU–A Assy (FRU 610). Refer to 6–2–78 on page 6–198.
6. Remove the Time Keeper RAM on the SBC–A Assy (MVME167–002B).
Note
After replacing a time keeper RAM with new one, it is necessary to install the system ID number into
the time keeper RAM and to set up the factory preset data.

Table of Contents

Other manuals for GE LOGIQ 400

Related product manuals