LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–206
REV 9
6–2–82 OMEM Assy (FRU No. 614)
Time Required
30 Minutes
Tool Required
Screwdriver
Procedure
CAUTION
Do not bend or flex the boards when mounting/dismounting the piggyback boards. Surface
mount IC boards are very susceptible to damage from flex/torque.
Refer to ILLUSTRATION 6–117 on page 6–207.
1. Turn OFF the system.
2. Remove the Side Cover Left (FRU 301). Refer to 6–2–31 on page 6–127.
3. Unscrew five Nest Cover screws.
4. Remove the Nest Cover.
5. Remove the TLMC Assy (FRU 606). Refer to 6–2–78 on page 6–198.
6. Unscrew one screw (
1
).
7. Remove the OMEM Assy out of the TLMC Assy.