GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–280
REV 9
6–7–4 Procedures (continued)
(A) Removing LV2 Unit (continued)
18. Disconnect one connector (
1
) from the rear of LV2 Unit. Refer to ILLUSTRATION 6–163.
19. Unscrew three screws (
2
–
4
) to remove the three wires. Refer to ILLUSTRATION 6–163.
20. Remove the LV2 Unit.
REAR VIEW OF LV2 UNIT
ILLUSTRATION 6–163
21. Disconnect the CN31 cable from the TRIAC Unit. Refer to ILLUSTRATION 6–164.
DISCONNECTING CN31 CABLE
ILLUSTRATION 6–164