GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
TM
8–14 OPTIONS
REV 0
8–2–7 Installing VCR on Color Monitor (Continued)
3. Remove the Monitor Cover Left and Monitor Cover Right. Refer to the 6–2–2 in Chapter 6, Renewal Parts.
4. Remove the left and right speakers. Refer to the 6–2–8 in Chapter 6, Renewal Parts.
5. Install the Plates to the Monitor at the both left and right side as shown in ILLUSTRATION 8–7. Use four screws
supplied with the VTR ON MONITOR ASSY (P9522YE) to fix the Plates.
6. Attach the Stopper Block on the Monitor Arm so as to cover the Up/down Release Button. Refer to ILLUSTRA-
TION 8–7.
PLATES AND STOPPER BLOCK INSTALLATION
ILLUSTRATION 8–7