Maintenance
2028451-182B MAC™ 1600 10-3
Inspecting the Equipment
Perform a visual inspection daily, preferably before the equipment’s first
use each day. During the inspection, verify that the device meets the
following minimum conditions:
The case and display screen are free of cracks and other damage.
All plugs, cords, cables, and connectors are free of kinks, frays, and
other damage.
All cords and connectors are securely seated.
All keys and controls operate properly.
If you notice any items that need repair, contact an authorized service
representative to make the repairs. Discontinue using the device until
the appropriate repairs can be made.
Cleaning the Device
Clean the exterior surface of the MAC 1600 device monthly, or more
frequently if needed.
Cleaning Materials to Use
Use the following materials to clean the device:
Mild dishwashing detergent
Clean, soft cloth (2)
Water
Cleaning Materials to Avoid
DO NOT use any of the following materials to clean the device, because
their use may damage equipment surfaces.
Organic solvents
Ammonia-based solvents
Abrasive cleaning agents
Alcohol
Virex
Sani-Master
Cleaning the MAC 1600 Surfaces
Use the following procedure to clean the surfaces of the MAC 1600
device.
1. Dilute mild dishwashing detergent in water to create a cleaning
solution.