10-12 MAC™ 1600 2028451-182B
Maintenance
discharges. To prolong the battery’s charge when not in use, store the
battery in a cool, dry location.
A new, fully-charged battery should last for approximately 3 hours of
normal operation. An on-screen LED indicates the condition and capacity
of the battery’s charge. (For more information on the battery gauge, refer
to “Front View” on page 2-2 and “System Errors” on page A-7). When the
LED flashes amber, connect the MAC 1600 system to AC power to charge
the battery to full capacity.
As the battery ages, the full charge capacity of the battery will degrade
and be permanently lost. As a result, the amount of charge that is stored
and available for use is reduced. When the capacity is no longer sufficient
for your daily operation, you will need to replace the battery.
Battery Safety
Observe the following warnings whenever handling the MAC 1600
battery.
WARNING
EXPLOSION OR FIRE — Using non-recommended
batteries could result in injury/burns to patients or users
and may void the warranty.
Use only batteries recommended or manufactured by GE.
WARNING
PHYSICAL INJURY — Leaks from battery cells can
occur under extreme conditions. The liquid is caustic to
eyes and skin.
If the liquid comes in contact with eyes, skin, or clothing,
flush with clean water and seek medical attention.
WARNING
BATTERY PACK DISPOSAL — Do NOT dispose of the
battery by fire or burning.
Follow local environmental guidelines concerning disposal
and recycling.
Replacing the Battery
When the battery’s full-charge capacity can no longer operate the MAC
1600 device for an adequate length of time, use the following instructions
to replace the battery.