10-14 MAC™ 1600 2028451-182B
Maintenance
For details, refer to “Replacing the Battery” on page 10-12.
2. Disconnect the AC mains power from the MAC 1600 unit.
3. Enter the Battery Status Service Diagnostic window.
4. Allow the battery to discharge until its Charge Level is less than
90%.
5. Turn off the unit and reconnect the AC mains power.
6. Allow the battery to fully charge.
The Battery LED will be solid amber while it is charging and turn
off when charging is complete.
7. Remove the AC mains power and turn on the MAC 1600 unit.
8. Allow the battery to discharge until MAC 1600 unit shuts down.
9. Reconnect the AC mains power to the MAC 1600 unit and leave the
unit turned off.
10. Allow the battery to fully charge.
When the Battery LED indicator stops flashing and turns solid, the
battery is fully charged and the conditioning cycle is complete.
Supplies and Accessories
For a list of available supplies and accessories for the MAC 1600 ECG
Analysis System, refer to the MAC 1600 ECG Analysis System Service
Manual (2028451-182).