10-4 MAC™ 1600 2028451-182B
Maintenance
2. Soak a clean cloth in the solution and wring out any excess.
3. Thoroughly wipe the surface of the MAC 1600 device with the damp
cloth.
Do NOT drip the solution or any liquid on the writer assembly.
Avoid contact with open vents, plugs, or connectors.
4. Repeat step 2 and step 3 as necessary until the surface is adequately
cleaned.
5. Wipe the surfaces with a dry, clean cloth or paper towel.
Cable and Leadwire Maintenance
Proper care and maintenance of the cables and leadwires used by the
MAC 1600 ECG Analysis System consists of:
Cleaning the cables and leadwires,
Storing the cables and leadwires, and
Replacing the cables and leadwires.
NOTE
The information in this section applies to the Multi-Link acquisition
cable and leadwires. For systems with the optional KISS system, see
the KISS operator’s manual for maintenance information.
Sanitizing Cables and Leadwires
Cables and leadwires come into contact with patients and, therefore,
should be cleaned and disinfected after every use. If necessary, they can
also be sterilized.
Before cleaning and disinfecting the cables and leadwires, you need to
know:
What cleaning materials can be used
What disinfectants can be used
What cleaning materials should be avoided
Cleaning Materials to Use
Use the following materials to clean the cables and leadwires:
Mild dishwashing detergent
Clean, soft cloth (2)
Water