A-4 MAC™ 1600 2028451-182B
Troubleshooting
ACI-TIPI Statement is Not Included on Report
If the ACI-TIPI statement does not appear when expected, do the
following:
Verify the ACI-TIPI option is activated.
For information on activating the ACI-TIPI option, refer to “Options
Setup” on page 9-29.
Verify ACI-TIPI is enabled on the ECG.
For information, refer to “Resting ECG Setup” on page 9-6.
Verify the ACI-TIPI required information was entered.
The ACI-TIPI statement will print only if the patient’s gender, date
of birth, and chest pain indication are included in the patient
information.
Verify the patient is 16 or older.
The ACI-TIPI statement will not print for pediatric patients.
Verify the original ECG was acquired in an electrocardiograph with
the ACI-TIPI option.
If you attempt to print an ECG that was imported from an external
device, the MAC 1600 device will not generate an ACI-TIPI
statement; it will print only if the statement was saved as part of the
ECG.
No BP Readings from Ergoline 900 Ergometer
If blood pressure readings were not taken when expected, do the
following:
Verify the device is supported.
For a list of supported devices, refer to “Connecting External Devices
(Stress Option)” on page 2-10.
Verify the device is connected to the MAC 1600 unit.
External stress devices connect to the MAC 1600 unit via a serial
cable. For more information, refer to “Back View” on page 2-3.
Verify the protocol is set up to take blood pressure.
For more information, refer to “Editing Stress Protocols” on
page 9-16.
Verify the device is set up correctly to acquire blood pressure.
Refer to the product documentation that accompanied the device.
External Device Does Not Move
If the external device does not move automatically when expected, do the
following:
Verify the correct device is selected in Basic Setup.
For more information, refer to “Basic Setup” on page 9-2.

Table of Contents

Other manuals for GE MAC 1600

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1600 and is the answer not in the manual?

GE MAC 1600 Specifications

General IconGeneral
BrandGE
ModelMAC 1600
CategoryMedical Equipment
LanguageEnglish

Related product manuals