8-58 Section 8-4 - System Modules- Replacement Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-4-10-5 MPB Module Installation Procedure
1.) Insert the MPB module as follows:
a.) Insert the MPB module so that it goes under the Top Receptical Box. To do that, the MPB
Module must be inserted in angled position and not at 90°.
b.) Make sure all cable connectors are floating.
c.) Position the MPB Module at 90° and slide into the rails until the docking of the MPB module.
2.) Tighten the two mounting screws to secure the MPB module.
3.) Return the PSU module.
4.) Return the battery packs.
5.) Refit the following covers: MPB Door Cover, Lower Front eTower cover, Right Side and Left Side
eTower covers, and RS Probe cover
6.) Install all accessories.
7.) Turn ON power to the system.
PSU Module Installation Procedure
Battery Module Installation Procedure
MPB Door Cover Installation Procedure
Left Side eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
Lower Front eTower Cover Removal Procedure
RS Probe Cover Installation Procedure
Accessories - Replacement Procedures