8-102 Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-6-8-5 MPB Fan Installation Procedure
1.) Carefully insert the new MPB fan into the fan sleeve.
2.) Tighten the elastic spring slings to secure the fan in its place.
3.) Reconnect the MPB to BIB Control Cable to the fan cable connector.
4) Install the MPB Front Metal Door.
Figure 8-87 Tightening Elastic Spring Slings
Figure 8-88 Installing the MPB Fan
MPB Front Metal Door Installation Procedure