Chapter 8 - Replacement Procedures 8-119
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
monitor, as shown below.
9.) Refit the cockpit upper and then the lower covers.
10.)Route the MPB to Cockpit cable through the cable cover door and the arm hole.
11.)Connect the MPB TO COCKPIT CABLE to the monitor:
Connect the cable connector (use flat-head screwdriver).
Connect Ground cable (use Phillips screwdriver).
Figure 8-109 Routing the MPB to Cockpit Cable under the Pin
Figure 8-110 Routing the MPB TO COCKPIT CABLE
Route cable under the pin