8-120 Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
12.)Measure 30 cm of cable between the connector and the arm.
13.)Secure cable to arm using tie wraps.
14.)Gently slide the MPB TO COCKPIT cable through the MPB receptacle side, towards the front.
15.)Reconnect J1 and J2 connectors ensuring the connectors' arrow is pointing upwards.
16.)Secure the cable to the plastic clips:
Three in the front end
Figure 8-111 Measuring 30 cm of the Cable
Figure 8-112 Securing the Cable to Arm
Figure 8-113 Reconnecting J1 and J2 Connectors