ADV100 • Functions description and parameters list 131
11074
(2)
(1)
(3)
(1)Setpoint
(2)Feedbackvalue
Example: suppose parameter 11002 Setpointvalof a pump pressure control is 10 bar, 11070 Constpressurethr
is set to 5%, 11074 Constpressoffsetis set to 4% and 11076 Fluidlossthr = 1% with 11078 Fluidlossthrtime
= 0.6 seconds
This means that to move to the constant pressure condition the feedback must be more than 9.5 bar.
We can consider 2 cases for restarting:
Case 1: the liquid leakage causes the pressure to fall by 0.1 bar per second
In this case the change is lower than the set threshold and to restart the drive the feedback must be less than 9.1
bar due to the settings in parameter 11074 Constpressoffset.
Case 2: the liquid leakage causes the pressure to fall by 0.2 bar per second
In this case the change exceeds that set in PAR 11076 Fluidlossthr(0.1 bar every 0.6 secongs). After 0.6 seconds
(from when the constant pressure threshold is exceeded) the pressure feedback is 9.38 bar, a leakage is detected
and the drive is restarted.
ALARMS
Menu PAR Description UM Type FB BIT Def Min Max Acc Mod
11050 Feedbacklosstime sec FLOAT 0.00 0 10000 RW FVS
This parameter denes the time for recognising the loss of the feedback signal.
The alarm is only recognised if parameter 11004 Feedbackselis set to PAR 1600 Analoginput1Xmonand PAR
1650 Analoginput2Xmon.
The drive recognises the alarm if the feedback signal is equal to zero (no signal) for longer than the time set in PAR
11050 Feedbacklosstime. If this parameter is set to zero, the alarm is not recognised.
Menu PAR Description UM Type FB BIT Def Min Max Acc Mod
11052 Feedbacklossact ENUM Stop 0 4 RW FVS
This parameter denes drive activity following the recognition of a PerditaretroazPID alarm.
0 Ignore
1 Warning
2 Disable
3 Stop
4 Faststop
Menu PAR Description UM Type FB BIT Def Min Max Acc Mod
11054 Feedbacklevel % FLOAT 20.000 1.000 50.000 RW FVS
This parameter denes the error level (expressed as a percentage with respect to the set point) for generating the
PIDFBloss alarm.
The alarm is generated when the PID error is higher than the set threshold.