Loading...
23W123W1
GENERAL INFORMATION
SERVICE AND REPAIR
TECHNICAL DRAWINGS
CONTROL
Guldmann develops, manufactures,
distributes and maintains products and
services that make the daily lives
of disabled people and their
carer a little easier.
SERVICE GUIDE
FOR GH
Vers. 3.00

Table of Contents

Questions and Answers

  • D
    David WalkerSep 8, 2025
    What to do if Guldmann Mobility Aid shows fault 18/22?
    • S
      shawncoxSep 8, 2025
      If your Guldmann Mobility Aid displays fault 18/22, indicating a hardware overcurrent on the leader of the twin hoist (18), you should consider: Replacing the motor (18). If the issue persists, replace the PCB (18).
  • A
    Alexis EvansSep 12, 2025
    What to do if Guldmann Mobility Aid is not charging 17?
    • N
      Nathan TerrySep 12, 2025
      If your Guldmann Mobility Aid is not charging (17), and the issue is with the follower charger, here’s what to check: * Check the connection between the two charger PCBs (J2 to J2). * Check the connection between the charger PCB (J4) and the motor safety PCB (J20 on 17). * Check the connection between the battery and charger (J3). If these steps don't resolve the issue, change the charger PCB.
  • D
    Deborah EdwardsSep 12, 2025
    What to do if Guldmann Mobility Aid shows fault 81/1?
    • K
      Kathleen LewisSep 12, 2025
      If your Guldmann Mobility Aid displays fault 81/1, indicating a communication fault with the Com module (81), you can try the following: * Check the cables and connection to the Com module (J105 or J5 on 81 to J2 on 16). * Check the jumper setting on (81). * If problems persist, replace (81).
  • A
    Andrea SalazarSep 14, 2025
    What to do if Guldmann Mobility Aid shows fault 82/1?
    • A
      angela75Sep 14, 2025
      If your Guldmann Mobility Aid displays fault 82/1, indicating a communication fault with the Com module follower twin hoist (82), you should: Check the cables and connection to the Com module (J105 or J5 on 82 to J2 on 18). Check the jumper setting on (82). If the problem persists, replace (82).
  • S
    Sophia WilliamsSep 16, 2025
    What to do if Guldmann Mobility Aid shows fault 32/1?
    • J
      James GreeneSep 16, 2025
      If your Guldmann Mobility Aid displays fault 32/1, indicating a communication fault with the Aux drive motor (32), here’s what to do: * Check the cables and connection to the aux module (J2 on RS485 to J3 on 16). * Check the Jumper setting on (32). * If the issue persists, replace (32).

Summary

General Information

General Conditions and Information

Basic terms, legal aspects, and general guidelines for use and handling.

GH Lifting Modules and Rail Types

Details on hoist models, configurations, and available rail system types.

Service and Repair

GH3 Service Procedures

Step-by-step instructions for servicing and repairing the GH3 hoist.

GH1 Service Procedures

Step-by-step instructions for servicing and repairing the GH1 hoist.

Technical Drawings

Spare Part Overviews

Illustrated lists of spare parts for various hoist models.

Wiring and Electrical Diagrams

Schematics for hoist configurations, including jumper settings.

Control and Inspection

Hoist and Rail System Inspection Guidelines

Procedures for periodic safety inspections according to standards.

Sling Inspection Guidance

Safety procedures for inspecting lifting slings.

Guldmann GH Specifications

General IconGeneral
TypeCeiling Hoist
Lifting Speed0.12 m/s
Max Speed0.12 m/s
Control TypeHandheld Control
Safety FeaturesEmergency stop, overload protection

Related product manuals