Section 6
Page 107
UVTOCOperation.fm Manual Control
6.12 Manual Control
The manual control menu allows manual control of the analyzer. It is entered
by pressing the UP button while the Calibration menu is displayed. The
manual control menu is then placed on the screen.
6.12.1 Initiate Clean
Press ENTER. The Initiate cln? screen is displayed.
Press ENTER to initiate a clean. Following a clean cycle the analyzer will
go online.
6.12.2 Initiate Calibration
Press ENTER to initiate an automatic calibration. A cal confirmation screen is
briefly displayed. After an auto-calibration the analyzer will go online.
6.12.3 Initiate Validation
Press ENTER to initiate an automatic validation. After an auto-validation the
analyzer will go online.
6.12.4 Go Online
Press ENTER to cancel a validation/calibration/clean cycle in progress.
A cal/cln cancelled confirmation screen is briefly displayed before going
through a purge to online.
Manual control
Initiate cln?
Initiate cal?
Initiate cal?
** Initiating **
Initiate val?
Go online?
Go online?
** Initiating **