5.2.2 Auto Options Screen
To enter this screen, press the UP button while in the Range select screen.
The display changes to:
Press ENTER. The Autocal standard screen appears.
5.2.2.1 Calibration After Validation
This screen enables or disables an auto calibration cycle following an
auto validation failure. Press ENTER to toggle between YES and NO.
The selection is saved upon exit from the screen.
5.2.2.2 Validation Tolerance
The Validation Tolerance screen allows the selection of the tolerance used in
Auto validation (refer to section 6.9 on page 95). To access this screen, press
the UP button while in the Auto validation screen.
5.2.2.3 Spill Threshold
If the value of TOC exceeds this % of the analyzer's range, the analyzer will
go offline for the selected spill recovery time (refer to section 5.2.2.4). This is
to allow the analyzer to return to online status as soon as the spill is rectified.
5.2.2.4 Spill Recovery
Press the DOWN button while in the Spill threshold screen. The Spill recovery
screen is displayed.
The Spill recovery is the time period that the analyzer will stay offline until a
spill, as detected in Spill threshold, is rectified. After this time period, it will
immediately go back online. The time period may be set to any number
between 0 and 99 minutes.
5.2.2.5 Spill Threshold
If the TOC level exceeds the spill threshold, the analyzer will clean for the spill
recovery time. Following the clean, it will purge and then go online to measure
the alternate stream.
Auto options
Cal after val
NO
Validation tol
5%
Spill threshold
100%
Spill recovery
20 min