Calibration and Configuration 111
Haemonetics
®
ACP-215
®
Service Manual P/N: SM-ACP215-01-EN(AB)
5. Press MODIFY PROGRAM until MARKET is selected and press
START. The device displays:
6. Press START. Choose between INTERNATIONAL and NORTH
AMERICA according to the geography of the install.
7. Press STOP three
times. The device displays:
8. Switch the memory protect switch located toward the bottom of the CPU
board (the left-most card cage board, looking from the rear of the unit) to
the Protect On mode (up). The device displays:
Note: The market configuration must be set prior to entering the protocol
configuration (see Configuration Parameter Default Settings tables)
MARKET CONFIGURATION
PLEASE SELECT MARKET
—> INTERNATIONAL
NORTH AMERICA
CONFIGURATION
—> PROTOCOLS CONFIGURATION
LANGUAGE
******************************
** REPLACE MEMORY**
** PROTECTION**
******************************
UTILITIES
CALIBRATION
DIAGNOSTICS
—> CONFIGURATION
MAINTENANCE
AUTO-CYCLE TEST