124 Calibration and Configuration
P/N: SM-ACP215-01-EN(AB) Haemonetics
®
ACP-215
®
Service Manual
Maintenance Procedures
Procedure # 1. Access the Utilities mode of operation by pressing and holding SHAKER
while switching the power on. Hold SHAKER until the UTLITIES menu
appears on the display.
2. Press MODIFY PROGRAM to move the arrow to MAINTENANCE and
then press START. The following display appears:
3. Press START then MODIFY PROGRAM to move the arrow to
PROCEDURE # and the following display appears:
4. Press START and the following display appears:
5. Press START and the following display appears:
MAINTENANCE
PRINT CALIB
—> PROCEDURE
CLOCK
EXECUTED PROCEDURES
#Executed Procedures XXXX
EXECUTED PROCEDURES
#Executed Procedures XXXX
PRESS START AGAIN TO RESET TO 0
OR PRESS STOP TO EXIT
EXECUTED PROCEDURES
#Executed Procedures 0