Preventive Maintenance and Installation 75
Haemonetics
®
ACP-215
®
Service Manual P/N: SM-ACP215-01-EN(AB)
Figure 19, Printer controls
3. On the UTILITIES menu, press MODIFY PROGRAM until the arrow
points to MAINTENANCE, then press START.
4. Press MODIFY PROGRAM until the arrow points to PRINT CALIB, then
press START.
5. Press MODIFY PROGRAM until the arrow points to PRINT CALIB
VALUES, then press Start.
6. Verify that the printer prints out the calibration values.
Functional/
Pump/Shaker
Check
Functional Test
1. Power on the device.
After completion of system tests and cover lock test, the screen displays:
HAEMONETICS ACP 215
PLEASE SELECT PROTOCOL
-> GLYCEROLIZATION
DE-GLYCEROLIZATION