Preventive Maintenance and Installation 87
Haemonetics
®
ACP-215
®
Service Manual P/N: SM-ACP215-01-EN(AB)
2. Plug the shaker in and press YES to continue. The screen displays:
The shaker will restart.
3. Once the shaker starts to shake, try to slow down the shaker using force
from yours hands. No errors should be displayed on the machine.
4. Press SHAKER and verify that the shaker has stopped.
5. Press SHAKER again to restart the shaker.
6. Use a Meylan, Model 109 Digital Tachometer (or equivalent) to measure
the operating speed (RPM). Verify that it is 180 +/- 2 RPM.
Rotor
Durometer
Test
Note: This test is performed during the annual preventive maintenance by
Haemonetics personnel only.
Procedure - Method 1
Materials Required
Hexagonal head wrench
TF9574-00 Roller Durometer Support
Rex Model H-1000-A Durometer Gauge
DE-GLYCEROLIZATION PROTOCOL
DILUTION 1 OF 2
DILUTING W/ WASH SOLUTION
PUMP TIME HYPE NACL ADDI
REAL: xx x 0 xxx 0
TARG: xx 56 50 1470 240