Preventive Maintenance and Installation 61
Haemonetics
®
ACP-215
®
Service Manual P/N: SM-ACP215-01-EN(AB)
General
Preventive
Maintenance
Frequency
General maintenance of the device should be performed as described in the
operation manual.
The preventive maintenance procedure outlined in this chapter should be
performed at intervals of 12 months +
30 days, although maintenance may be
required more often if the units receive frequent use, with the following
exceptions:
Any time the device is disassembled to any extent beyond removing the
access panels.
Any time there is a fluid spill on the device or other potential physical
damage that may have affected the integrity of the unit.