118 Calibration and Configuration
P/N: SM-ACP215-01-EN(AB) Haemonetics
®
ACP-215
®
Service Manual
Language 1. Access the Utilities mode of operation by pressing and holding SHAKER
while switching the power on. Hold SHAKER until the UTLITIES menu
appears on the display.
2. Press MODIFY PROGRAM to move the arrow to CONFIGURATION and
then press START.
3. The message REMOVE MEMORY PROTECTION OR STRIKE STOP
appears.
4. Switch the memory protect switch located toward the bottom of the CPU
board (the left-most card cage board, looking from the rear of the unit) to
the Protect Off mode (down). The calibration screen appears with three
choices: PROTOCOLS CONFIGURATION, LANGUAGE, and
MACHINE.
5. Press MODIFY PROGRAM to move the arrow to LANGUAGE, then
press START.
6. Press START. The following screen is displayed:
7. Press MODIFY PROGRAM to move the arrow to ENGLISH, or desired
language, then press START. The following screen is displayed:
8. Press the STOP key two times to exit. The following message appears:
LANGUAGE SELECTION
—> ENGLISH ITALIAN
FRENCH DUTCH
GERMAN
Language = XXXXXXX
LANGUAGE SELECTION
—> ENGLISH ITALIAN
FRENCH DUTCH
GERMAN
Language = ENGLISH
CONFIGURATION
PROTOCOLS CONFIGURATION
—> LANGUAGE
******************************
** REPLACE MEMORY**
** PROTECTION**
******************************

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Haemonetics ACP 215 and is the answer not in the manual?

Haemonetics ACP 215 Specifications

General IconGeneral
BrandHaemonetics
ModelACP 215
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals