86 Preventive Maintenance and Installation
P/N: SM-ACP215-01-EN(AB) Haemonetics
®
ACP-215
®
Service Manual
Once the HYPE REAL volume reaches 50 the pumps and shaker stop
and the screen displays:
After the delay the screen displays:
The BLOOD pump, SOLUTION pump, and the SHAKER will start.
Shaker Connection/Speed Test
1. Unplug the shaker.
After a time the screen displays:
The device emits a beeping alert sound.
DE-GLYCEROLIZATION PROTOCOL
HYPERTONIC EQUILIBRATION DELAY
PUMP TIME HYPE NACL ADDI
REAL: x x 0 0 0
TARG: x 56 50 1470 240
DE-GLYCEROLIZATION PROTOCOL
DILUTION 1 OF 2
DILUTING W/ WASH SOLUTION
PUMP TIME HYPE NACL ADDI
REAL: x x 0 0 0
TARG: x 56 50 1470 240
DE-GLYCEROLIZATION PROTOCOL
Check Shaker Connection
Press YES to continue
PUMP TIME HYPE NACL ADDI
REAL: x x 0 0 0
TARG: x 56 50 1470 240