Preventive Maintenance and Installation 85
Haemonetics
®
ACP-215
®
Service Manual P/N: SM-ACP215-01-EN(AB)
4. If the following screen appears, press NO to continue until the blood
pump speed test (CW) is done:
5. Apply a negative pressure to the DPM.
6. Set meter to mmHg scale. Once the pressure on the pressure meter
reads –200 mmHg. The screen displays:
Once this is done the screen displays:
The BLOOD pump SOLUTION pump and the SHAKER will start.
7. Quickly measure the speed of the Blood and the Solution Pumps. They
will be running at 150 +/- 2 RPM. Record these values.
- IS FINAL PRODUCT BAG EMPTY?
Press YES to start Prime
Press NO to continue Purge
PUMP TIME HYPE NACL ADDI
REAL: x x 0 0 0
TARG: x 56 50 1470 240
DE-GLYCEROLIZATION PROTOCOL
PRIMING HYPERTONIC SOLUTION
PUMP TIME HYPE NACL ADDI
REAL: x x 0 0 0
TARG: x 56 50 1470 240
DE-GLYCEROLIZATION PROTOCOL
DILUTING W/ HYPERTONIC SOLUTION
PUMP TIME HYPE NACL ADDI
REAL: x x xx 0 0
TARG: x 56 50 1470 240