84 Preventive Maintenance and Installation
P/N: SM-ACP215-01-EN(AB) Haemonetics
®
ACP-215
®
Service Manual
32.Press YES to continue. The screen displays:
Press START to begin the procedure. The screen displays:
Blood Pump Speed Test (CW)
The blood pump rotates clockwise at 100 RPM.
1. Ensure that the displayed pump speed is 100 ml/min +/- 2 ml/min.
2. Use a contact type digital tachometer, Shimpo Model DT105 (or
equivalent) with the tip in contact with the pump rotor mounting screw to
measure the pump speed in RPM.
3. Measure the speed of the BLOOD pump and verify the reading is 100 +/-
2 rpm.
DE-GLYCEROLIZATION PROTOCOL
PRESS MODIFY TO SET PARAMETERS
PRESS START TO BEGIN
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
DE-GLYCEROLIZATION PROTOCOL
PURGING AIR FROM PRODUCT BAG
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240