Troubleshooting 155
Haemonetics
®
ACP-215
®
Service Manual P/N: SM-ACP215-01-EN(AB)
5 ROM TEST FAILURE ROM CRC does not
match value stored in
NVRAM.
1. Replace ROM IC (Main program IC).
2. Replace CPU PCB.
7 CALIBRATION CRC
FAILURE
Calibration CRC does not
match value stored in
NVRAM.
1. Recalibrate the A/D, and centrifuge.
2. Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
centrifuge and configure the language.
Reset protocol parameters (MODIFY
PROGRAM) if parameters were
different from default values.
10 DPM FAILED 20%
ANALOG TEST
DPM channel of A/D
converter failed 20%
analog voltage load test.
Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
CENTRIFUGE, and configure the
language. Reset protocol parameters
(MODIFY PROGRAM) if parameters
were different from the default values.
10 DPM FAILED 50%
ANALOG TEST
DPM channel of A/D
converter failed 50%
analog voltage load test.
Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
CENTRIFUGE, and configure the
language. Reset protocol parameters
(MODIFY PROGRAM) if parameters
were different from the default values.
10 DPM FAILED 80%
ANALOG TEST
DPM channel of A/D
converter failed 80%
analog voltage load test.
Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
CENTRIFUGE, and configure the
language. Reset protocol parameters
(MODIFY PROGRAM) if parameters
were different from the default values.
11 SPM FAILED 20%
ANALOG TEST
SPM channel of A/D
converter failed 20%
analog voltage load test.
Replace CPU PCB. Using Utilities,
calibrate all components under A/D, and
CENTRIFUGE, and configure the
language. Reset protocol parameters
(MODIFY PROGRAM) if parameters
were different from the default values.
Table 9, CPU error messages
Cod
e
Text Message Description Suggested Resolution

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Haemonetics ACP 215 and is the answer not in the manual?

Haemonetics ACP 215 Specifications

General IconGeneral
BrandHaemonetics
ModelACP 215
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals