Principles of Operation 55
Haemonetics
®
ACP-215
®
Service Manual P/N: SM-ACP215-01-EN(AB)
Note: These parameters are market dependent and are only changed if the
“Machine Market” configuration is changed.
The following flow chart detail the User interface within the protocol
configuration of the ACP215. These are market dependent and will be
available once the market configuration has been set under the machine
configuration
menu.
Note: See configuration setup procedure in Chapter 5.
Table 4, Configuration parameter default settings
International market North America market
Config parameter Default value Config parameter Default value
*Disposable Checklist Minimum Set *Disposable Checklist Full Set
*Modifiable Parameters Full Set *Modifiable Parameters Minimum Set
Available Protocols Frozen Blood Available Protocols Frozen Blood
Cell Wash Additive Option Enabled Cell Wash Additive Option Disabled
Bowl Size Option Small Bowl
Only (275ml)
Bowl Size Option Small Bowl
Only (275ml)
*Sterile Check in Cell Wash Enabled *Sterile Check in Cell Wash Disabled