EasyManua.ls Logo

HemoCue 201 DM - Default Settings

Default Icon
203 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
98
3.5 Default Settings
FIGURE 3-82
When the Default Settings button is pressed
in the Settings menu (see FIGURE 3-4),
confirmation or cancellation of the
requested change to the Analyzer default
setting is required. A question will be
displayed.
Default settings aect only the Analyzer
configuration (including password).
Measurement results or lists (Liquid Controls,
Cuvette Batches, Operators, Comments or
Logs) will not be aected. For a list of default
values see Appendix A Default Settings.
a) If all the settings in the Analyzer are to be
changed to Default settings, press
Continue.
Otherwise, press No. FIGURE 3-4 will be
displayed.
If the choice was to change all settings to
default, the following text will be displayed
for a few seconds:
Settings have Now Changed to Default
Settings.
b) The Settings menu, FIGURE 3-4, will be
displayed again.
Default
Settings
NoNo
ContinueContinue
Are You Sure You
Want to Change to
Default Settings?
All current settings
will be changed

Table of Contents

Other manuals for HemoCue 201 DM

Related product manuals