125
6 QC Test procedure
This chapter describes the process of performing QC Tests, Linearity Tests and
Proficiency Tests.
6.1 Events necessitating
QC Tests
6.1.1 QC Reminder
Based on the settings made for the QC Test
Reminder, the Reminder icon(1) will be
displayed to warn of an impending QC
lockout.
FIGURE 6-1
The QC Reminder(1) indicates Time(2) or
Number of measurements(3) remaining
before the QC Lockout.
The number of measurements remaining
refer only to Patient Tests and STAT tests.
The text inside the brackets indicates the
types of QC Tests(4) that must be performed
in order to avoid a QC lockout.
QC1 = QC Test Level 1
QC2 = QC Test Level 2
QC3 = QC Test Level 3
QC4 = QC Test Level 4
QC5 = QC Test Level 5
1 QC Reminder
2 Time remaining before QC Lockout
3 Number of measurements remaining before
QC Lockout
4 Type of QC Test required
08/09/03
10:55
AM
JAMES SMITH
2h 47min (QC1)
100 cuv (QC2)
1
3
2
4