131
6.2.1 QC Tests at a specific
concentration level
FIGURE 6-7
The QC Tests procedure for Level 1, Level 2,
Level 3, Level 4 and Level 5 are identical, and
therefore only one, Level 1, is described.
The concentration level of the QC Test(1) is
displayed.
a) Fill a Cuvette with the appropriate level of
Liquid Control(1).
b) Place the Cuvette in the Cuvette holder
and gently insert it into the measuring
position.
FIGURE 6-8 will be displayed.
1 Concentration level of QC Test
Please Fill
and
Insert a Cuvette
QC Level 1
1