EasyManua.ls Logo

HemoCue 201 DM - Verify;Duplicate Sampling

Default Icon
203 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
112
4.1.2 Verify/Duplicate sampling
Patient Test
PID: 750629-3694
LID: 12345
63 mg/dL
Result
1
FIGURE 4-13
Results may be verified by obtaining a
second sample from the same patient and
performing a new measurement. Depending
on whether the Duplicate sampling function
(see Duplicate sampling) has been activated
or not the outcome of this procedure may
vary.
Never re-measure a Cuvette!
a) When the first result is displayed, press
the Verify/Duplicate Sampling button(1).
FIGURE 4-14 will be displayed.
1 Verify/Duplicate sampling button
Patient Test
PID: 750629 -3694
LID: 12345
Cuv.123456
Verify
Press "OK"
to Accept
2
FIGURE 4-14
b) Accept the information for the new
measurement by pressing the Confirm
button(2).
c) The following text will be displayed:
Please Fill and Insert a Cuvette.
d) When the new Cuvette is inserted and the
measurement is in progress, the following
text will be displayed:
Please Wait Measuring…
e) Upon completion of the measurement,
the result is displayed, see FIGURE 4-15.
2 Confirm button

Table of Contents

Other manuals for HemoCue 201 DM

Related product manuals