Symptom Explanation Action
Overrange HemoCue Hb 201 DM:
Measured value exceeds 25.6 g/
dL (256 g/L, 15.9 mmol/L).
HemoCue Glucose 201 DM:
Whole blood:
The measured value exceeds 400
mg/dL (22.2 mmol/L).
Plasma Equivalent:
The measured value exceeds 444
mg/dL (24.6 mmol/L).
HemoCue Glucose 201 DM RT:
Whole blood:
The measured value exceeds
500mg/dL (27.8mmol/L).
Plasma Equivalent:
The measured value exceeds
560mg/dL (31mmol/L).
HemoCue Hb 201 DM and HemoCue
Glucose 201 DM RT:
The result is above the measuring
range.
HemoCue Glucose 201 DM:
Whole blood:
The measuring range may be
extended to 800 mg/dL (44.4
mmol/L) by using the Dilution
function.
Plasma Equivalent:
The measuring range may be
extended to 888 mg/dL (49.2
mmol/L) by using the Dilution
function.
Note that the Dilution function is not
available in all markets.
Non-
responsive
display
Display needs recalibration. Follow section 5.4 Calibrating the
Display in Instructions for Use.
If the recalibration fails, the Analyzer
needs service. Contact the
distributor.