CPR Control
Progressa™ Bed User Manual (171528 REV 1) 13
CPR Control
The red CPR control pedals are located on each side of the base frame between the head-end and
foot-end casters.
Use the CPR Control
When connected to AC power, the HandsFree® CPR Control
lowers the head and knee sections, and raises the foot section.
After the head section is flat, a tone sounds and the foot section
rises. The foot section moves to a flat position within a maximum
of 30 seconds if fully articulated.
The integrated air surface will Max-Inflate to provide a firm
surface to support a CPR board. After 60 minutes of
Max-Inflate, the optional air surface will go into Normal mode. If AC power is lost, the air surface stays
at the level of pressure that existed at the time of power loss.
To Activate:
1. Step down and hold the red CPR pedal with your foot until the head section reaches the flat
position and you hear the audible tone. If you release the CPR pedal before the bed is flat, the
head section will stop.
– The foot and knee sections will automatically move to a flat position from any position
including chair.
2. The surface automatically goes into Max-inflate for 60 minutes. After 60 minutes the surface
will go into Normal/Standard mode.
NOTE:
Use of a CPR board may increase the effectiveness of CPR.
3. To stop foot section movement, press any other siderail control except Nurse Call.
4. To stop Max-Inflate, press the Surface control on the GCI home screen.
Then press Normal.
NOTE:
When the AC power is lost, the head section will lower and the foot
section will raise. The optional integrated air surface will not Max-Inflate
and CPR board effectiveness may be reduced.
The Bed Up/Down controls are usable when the CPR function is activated.
When CPR is activated, any controls that are locked out will become unlocked.