Caregiver Siderail Controls
Progressa™ Bed User Manual (171528 REV 1) 25
Bed Flat Control
Bed Flat controls are provided so that a caregiver can easily return the patient deck to
the level position from any articulated position.
1. Press and hold the Bed Flat control.
2. The intermediate frame returns to level from the tilted position.
3. The individual sections move to the flat position. If the bed is starting in the Chair position, then
it will move through the Reclining position on the way to level.
4. When all sections are flat, the bed stops and one beep sounds.
Nurse Call
Use the Nurse Call control to activate the Nurse Call feature. The controls are located on both the
inboard and outboard sides of the intermediate siderails. The Nurse Call controls are always active if
connected to the facility communication system.
To Activate:
Press a Nurse Call control.
When the nurse station acknowledges the nurse call, the inboard
indicator continuously illuminates amber and the outboard indicator does
not illuminate.
When the nurse station communication line is open, both the inboard
and the outboard indicators continuously illuminate green.
Speak into the speaker/microphone located on the inboard side of the
head end siderails.
Max-Inflate (Siderail Method)
To Activate
Press the Max-Inflate control.
The green indicator light will turn on.
To Deactivate
Press the Max-Inflate control
The green indicator light will turn off.
See "Max-Inflate" in the GCI section for an alternative method. This feature times out after 30 minutes.
Side Exit Assist
The Side Exit Assist control inflates the seat section of the surface to assist in side
exit of the bed. This feature times out after 30 minutes.
To Activate
1. Help patient to side sitting position on edge of mattress.
2. Raise or lower the bed so the patient's feet will be flat on the floor.
3. Press the Side Exit Assist control on the head end siderail.
4. After the seat section inflates, assist the patient with bed exit.
To Deactivate
Press the Side Exit Assist control on the head end siderail.