Caregiver Pendant Controls
26 Progressa™ Bed User Manual (171528 REV 1)
Caregiver Pendant Controls
This section describes the pendant controls that are intended to be used only by the caregiver.
WARNING:
The caregiver pendant is for use by the caregiver only. Store out of reach of the patient when not in
use. Patient injury can result.
WARNING:
The pendant is not for use inside of an oxygen tent. Patient injury can result.
There is an Enable control located on the caregiver pendant. The Enable control deters unauthorized
operation of controls on the pendant. The Enable control is only required for the functions connected
with the green line. The controls in the blue area do not require the Enable control to be activated.
The Enable indicator stays on for 60 seconds.
While this indicator light is on, the caregiver
can activate any pendant control.
To Activate
• Press and hold the Enable control until
the indicator light comes on. The
Enable indicator light stays on for 60
seconds.
• During the 60-second period, you may
activate bed controls on the pendant
without pressing the Enable control
again.
• If the enable control process is done
incorrectly, the bed will beep three
times and instructions will show on the
GCI.
• If during the 60 second enabled time
you want to turn it off, press the Enable
control. The indicator light will turn off
when the Pendant controls are no longer enabled.
To Remove from the Siderail or Footboard
• Pull straight up on the pendant.
or
• Rotate the pendant in a clockwise or counterclockwise direction until the mount clip
disengages from the siderail or footboard.
To Store
• Push straight down on the pendant until the mount clip engages on the siderail.
WARNING:
The pendant should not be stored under the mattress. Patient injury or equipment damage could
occur.