Surfaces
Progressa™ Bed User Manual (171528 REV 1) 57
Seat Deflate
The Seat Deflate feature allows for
easier bedpan placement.
WARNING:
Seat deflate is not
recommended for side sitting
or side egress.
To Activate
1. Press the Surface control
on the GCI.
2. Press the Seat Deflate
control.
The Therapy Surface will
automatically exit the Seat Deflate
mode and return to normal mode
after 30 minutes. After 28 minutes,
a beep will sound and a screen on
the GCI will appear that there are 2
minutes left.
To Deactivate
1. Press the Surface control
on the GCI.
2. Press the Normal control.
Patient Comfort
Allows customizing based on
patient request while maintaining
pressure redistribution.
The system automatically supplies
pressure distribution for the
patient’s position on the mattress.
To Adjust the Firmness
1. Press the Surface control
on the GCI.
2. Press Patient Comfort on
the GCI.
3. Use the Patient Comfort
controls to change the
pressure in the head, seat,
and lower leg sections of
the mattress assembly:
•To Increase the
pressure, press the
Up arrow.
•To Decrease the
pressure, press the
Down arrow.