97
────────────────────────────────────────────────────
7.4 Command
────────────────────────────────────────────────────
Startsatest.
Syntax
:STAR
Response
OK
Function
Starts a test in the READY state
In the optional functions, if "START" for the RS command is set to "0: Not set," this
command will not start testing.
Example
PC
>
3158
>
:STAR
OK Completed.
Error
The execution of this command in a state other than the READY state causes an
execution error.
In the optional functions, if "START" for the RS command is set to "0: Not set," an
execution error will occur.
Note
・
The ":STAR" command cannot be accepted for approximately 0.5 seconds after the
":STOP" command has been sent.
・
In this command, set Momentary Out in the optional function setting to "0:OFF".
At "1:ON", the test forcibly stops immediately after test starts.
Performs forcible ending of a test and releases the hold state.
Syntax
:STOP
Response
OK
Function
In the TEST state, performs forcible ending of a test.
When retaining the test result, returns to the READY state.
However, in the optional functions, if "FAIL Mode Function" is set to ON, the Hold
function cannot be disabled by this command.
Example
PC
>
3158
>
:STOP
OK Ended.
Error
In the optional functions, if "FAIL Mode Function" is set to ON, an execution will
occur when the unit is in the FAIL Hold Mode.
Note
The ":STAR" command cannot be accepted for approximately 0.5 seconds after the
":STOP" command has been sent.
:STAR