8.4 Displaying Errors
────────────────────────────────────────────────────
1
2
3
4
5
6
8
10
11
12
13
14
A
Err0 Interlock state. (See Section 6.1.4 )
Err1 There is a problem with the external switch. A device other
than a remote-control box (9613/9614) is connected. If this
error appears when a proper remote-control box is
connected, there may be a malfunction.
Contact your sales agent or nearest sales office.
Err2 The EXT I/O may be malfunctioning.
Contact your sales agent or nearest sales office.
Err3 The EEP-ROM may be malfunctioning.
Contact your sales agent or nearest sales office.
Err4 The RAM, ROM may be malfunctioning.
Contact your sales agent or nearest sales office.
Comparative voltage value setting OFF
Comparative voltage value 0.00 kV
Upper level value 0.2 mA
Lower level value setting OFF
Lower level value 0.1 mA
Test time setting ON
Test time 0.5 s
8.4 Displaying Errors
8.5 Resetting the System
box and packing materials. However, do not use if the box is torn or out of shape,
or if the packing materials are crushed.
・
When packing the instrument, make sure to remove the test leads and power
supply cords from the main device.
・
When transporting, protect it from strong impact such as dropping it.
If an error occurs, the 3158 displays the following on the screen.
Resetting the system
While pressing the
SHIFT
key, press the main power switch to turn on power.
Parameters after resetting the system
Once the system is reset, the following parameters are initialized together with the
saved setting data. The optional settings are reset to default values (initial settings).