(5) Double Action
Enable this function to allow testing to start only when the
START
key is pressed
within about 0.5 seconds after the
STOP
key.
Selection
0: Not set (initial setting) 1: Set
(6) FAIL Mode
Enable this function to restrict hold release to the
STOP
key on the main unit.
Selection
0: Not set (initial setting) 1: Set
(7) RS Command [START]
Turn this function on to enable the RS-232C START command.
Selection
0: Not set (initial setting) 1: Set
(8) Inter-lock Function
Enable this function to activate the external I/O interlock terminals.
Selection
0: Not set (initial setting) 1: Set
(9) Voltage Comparator Position
When the voltage comparator is set to ON for a withstand voltage test, select
whether you want to use the voltage comparator when starting and during the test,
or only when terminating the test.
Selection
0: Start test (initial setting) 1: End test