8.5 Message List
125
8
Chapter 8 RS-232C/GP-IB/LAN Interfaces
:SYSTem:EOM:PULSe
<HOLD/PULSe> Selects the EOM pulse width
151
:SYSTem:EOM:PULSe?
(0.001 to 0.099) Queries the EOM pulse width setting
151
Measurement current pulse output function setting
:SYSTem:CURRent
CONTinuous/PULSe
Sets the measurement current pulse
output function
151
:SYSTem:CURRent?
CONTINUOUS/PULSE Queries the measurement current
pulse output function
151
Terminator
:SYSTem:TERMinator
0/ 1 Sets the terminator 150
:SYSTem:TERMinator?
0/ 1 Queries the terminator 150
System Reset
:SYSTem:RESet
Executes a system reset, including
saved measurement setting state data
151
EXT I/O
:IO:OUT
0 to 1023 EXT I/O output 152
:IO:IN?
0 to 31 EXT I/O input 152
Trigger
:INITiate:CONTinuous
1/ 0/ ON/ OFF Sets continuous measurement 155
:INITiate:CONTinuous?
ON/ OFF Queries the continuous measurement
setting
155
:INITiate[:IMMediate]
Trigger wait setting 155
Trigger Source Setting
:TRIGger:SOURce
IMMediate/ EXTernal Sets the trigger source 156
:TRIGger:SOURce?
IMMEDIATE/ EXTERNAL
Queries the trigger source setting 156
:TRIGger:DELay:STATe
1/ 0/ ON/ OFF Sets the trigger delay 156
:TRIGger:DELay:STATe?
ON/ OFF Queries the trigger delay setting 156
:TRIGger:DELay
<Delay time> Sets trigger delay time 157
:TRIGger:DELay?
0 to 9.999 Queries the trigger delay time 157
Reading Measured Values
:FETCh?
<Resistance mea-
sured value>, <Voltage
measured value>
V mode
<Resistance mea-
sured value> mode
<Voltage measured
value> V mode
Reads the most recent measurement 157
:READ?
<Resistance mea-
sured value>, <Voltage
measured value>
V mode
<Resistance mea-
sured value> mode
<Voltage measured
value> V mode
Executes a measurement and read
the measured values
158
Message ([ ] = optional) Data Contents
( ) = response data
Description Ref
page