9.3 General Specifications
188
9.3 General Specifications
Operating temperature and
humidity
0 to 40°C (32 ± 104°F), 80% RH or less (non-condensating)
Storage temperature and humidity -10 to 50°C (14 ± 122°F), 80% RH or less (non-condensating)
Temperature and humidity range
for guaranteed accuracy
23 ± 5°C (73 ± 9°F), 80% RH or less (non-condensating)
Period of guaranteed accuracy 1 year
Operating environment Indoors, pollution degree 2, up to 2000 m (6562 ft) ASL
Rated supply voltage 100 V AC to 240 V AC (Figures reflect assumed voltage fluctuations of
10%.) (Auto selecting)
Anticipated transient overvoltage: 2500 V
Rated supply frequency 50 Hz/ 60 Hz
Power consumption BT3562, BT3562-01, BT3563, BT3563-01: 30 VA
BT3561A, BT3562A, BT3563A: 35 VA
Dielectric strength 1.62 kV AC for 1 minute, Cutoff current 10 mA, between all power
terminals and protective ground
3.00 kV AC for 1 minute, Cutoff current 1 mA, between all measure-
ment terminals and Interfaces
1.62 kV AC for 1 minute, Cutoff current 1 mA, between all measure-
ment terminals and protective ground
Dimensions Approx. 215W × 80H × 295D mm (8.46”W × 3.15”H × 11.61”D)
(excluding protrusions)
Mass Approx. 2.4 kg (84.7 oz.)
Accessories See: "Verifying Package Contents" (p.2)
Options See: "Options" (p.2)
Applicable Standards Safety EN61010
EMC EN61326 Class A
Effect of radiated radio-frequency
electromagnetic field
Resistance measurement : ± 10% of reading ± 8000 digits at 10 V/m
Voltage measurement :
± 0.01% of reading ± 50 digits at 10 V/m
Effect of conducted radio-frequency
electromagnetic field
Resistance measurement : ± 0.5%
of reading
± 1000 digits at 3 V
Resistance measurement : ± 0.5% of reading ± 1000 digits at 10 V
(BT3561A, BT3562A, and BT3563A only)
Product warranty period 3 years