3.5 Setting Sampling Rate
34
The sampling rate can be selected from EX.FAST, FAST, MEDIUM and SLOW.
Slower sampling rates generally provide greater measurement precision.
3.5 Setting Sampling Rate
Selects the sampling rate
EX.FAST FAST MEDIUM SLOW
• Measurements are especially susceptible to interference from the environ-
ment when EX.FAST is selected, so countermeasures such as shielding or
twisting of test leads, cables and wiring around the test object may be nec-
essary.
See "Appendix 1 Precautions for Making Custom Test Leads" (p.1)
• When SLOW sampling is selected, self-calibration is executed during each
measurement. At other sampling rates, self-calibration is executed manually
or automatically every 30 minutes.
See "4.10 Self-Calibration" (p.73).
• Refer to the specifications for details of sampling rates.
See " Sampling Time" (p.182).