9.1 Basic Specifications
183
9
Chapter 9 Specifications
* When SLOW sampling is selected, self-calibration is performed upon each measurement. In this state, the
calibration mode setting is ignored.
* When using the instrument with an external trigger source or with continuous measurement set to "Off,"
application of the measurement current can be limited to when measurement is being performed.
(This functionality is designed to prevent interference caused by measurement currents when using multiple
instruments simultaneously.)
Zero-Adjustment
Zero-adjustment function Zero-adjustment setting
ON/ OFF (Common to both resistance and voltage)
Zero-adjustment clear
Turns zero-adjustment off and clears all zero-adjustment offset data
Zero-adjustment range Resistance measurement : -1000 to 1000 counts
Voltage measurement : -1000 to 1000 counts
Self-Calibration
Calibration mode AUTO/ MANUAL
AUTO Executes automatically once every 30 minutes
MANUAL Executes manually by EXT I/O signal or remote command
Trigger
Trigger source Internal/ External
Measurement current control
Measurement current mode Pulse/ Continuous
Delay
Delay function ON/ OFF
Delay time 0 to 9.999 sec
Averaging
Averaging function ON/ OFF
No. of samples to average 2 to 16 times
Averaging method Moving average with internal triggering, and simple average with external
triggering
Comparator
Comparator function ON/ OFF (Common to both resistance and voltage)