9.4 Functions
171
9
Chapter 9 Specifications
Measurement memory
and batch download
functions
Measurement memory setting: ON/OFF/Clear
Memory trigger: Up to 400 measurement values can be stored in internal
memory by EXT I/O signals, key or remote command.
Stored measurement values can be batch downloaded by remote command.
* Data stored in memory cannot be displayed on the instrument.
Measurement value output function: Outputs measured values via the RS-
232C interface upon triggering
Key-lock function Key-lock setting: ON/OFF
Key operations are disabled when ON.
Power supply frequency
setting function
Operating power supply frequency setting: AUTO (automatic selection of 50
Hz/60 Hz)/50 Hz/60 Hz
Panel save function No. of panel to save: 126
Saved settings: Functions, resistance measurement range, voltage measure-
ment range, auto-ranging setting, zero-adjust on/off setting and value, sam-
pling rate, trigger source, delay setting, averaging setting, comparator setting,
statistical calculation setting, display toggle, key-lock setting
* Measurement conditions can be saved and loaded by specifying a panel
number.
Reset Reset method: Reset/System reset
* System Reset also initializes the panel save data
Display device LED
HIOKI BT3564A961-02