EasyManua.ls Logo

Hioki BT3564 - Comparator Setting Example 2 (Reference Value and Tolerance Judgment)

Hioki BT3564
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.1 Comparator Function
45
4
Chapter 4 Applied Measurement
This example describes the comparator setting method.
The upper and lower thresholds are saved as the displayed counts (indepen-
dent of measurement mode and range). Therefore, changing the measure-
ment mode or range results in the same display counts representing different
absolute values.
Example:
To specify the lower threshold as 150 m
in the 300 m range, enter “15000”.
Switching to the 3
range after making this setting changes the lower thresh-
old to 1.5
.
The instrument can also base judgments on the absolute value of measured
voltage values (to prevent Lo judgments when the positive and negative termi-
nals are connected backwards).
See "Configuring the Absolute Value Judgment Function (Voltage)" ( p.53)
Comparator Setting Example 2 (Reference Value and Tolerance Judgment)
Example:
Set a reference value and tolerance in the V mode (3 range), and set the beeper
to sound while measured values are within tolerance.
Resistance : Reference value 1.5 , Tolerance 5%
Voltage : Reference value 4.2 V, Tolerance 0.5%
1
Confirm that the Comparator function is off.
(The settings cannot be changed while the Comparator function is enabled. Press the
COMP key to disable the Comparator function.)
2
Select the V measurement mode.
3
Select the measurement range (for this example, the 3 range).
COMP not lit
V lit
Increase the resistance measurement range.
Decrease the resistance measurement range.
lit
HIOKI BT3564A961-02

Table of Contents

Related product manuals