4.1 Comparator Function
44
12
Press so that the indicated position blinks, and select the comparison method for
the comparator (for this example, HIGH/LOW).
HIGH, LOW ..... Compare by upper and lower thresholds (default setting)
REF, %............. Compare by reference value and tolerance
13
Switch to the upper/lower threshold setting display, and specify the thresholds.
14
Applies the settings and returns to the measurement display.
The comparator function is enabled.
To cancel the settings: SHIFT key
15
Connect a test object and judge the measured value.
In the V mode, you can verify comparator settings by pressing the VIEW key.
See
" Switching Between Measurement Value and Comparator Setting Displays" (
p.56).
HIGH and LOW flashing
Or
numeric key-
pads
For this example,
Upper threshold: 15.2 V
Upper threshold: 15 V
COMP lit
Measured resistance value
Judgment result
Measured voltage value
Upper threshold value < Measured value
Lower threshold value
Measured value Upper threshold value
Measured value < Lower threshold value
HIOKI BT3564A961-02