3.6 Zero-Adjust Function
34
Clearing Zero-Adjustment
1
(The SHIFT indicator lights up.)
The Menu display appears.
2
The zero-adjust value function is canceled. (0ADJ not lit)
(Main display)
(Sub display) flashing
(Main display)
If Err02 is displayed
Indicates that zero adjustment could not be executed, either because the range
to be adjusted exceeds
± 1000 dgt, or a measurement fault condition exists.
The zero adjust function is canceled, so repeat the operation after correcting
the cause of the error.
Zero adjustment is limited to ± 1000 dgt. (all ranges)
Perform zero adjustment for each range that is used in measurement.
When using the auto-range function, perform zero adjustment for all ranges.
When using the
V function, the 0ADJ indicator lights up or turns off
according to the resistance measurement range zero-adjust state.
Zero-adjustment values are retained even when power is turned off.
The 0ADJ terminal of the EXT I/O connector also executes zero adjustment.
See "5.2 Signal Descriptions" ( p.76).